- FDA identifier
- Z-0085-2014
- Company in FDA record
- Baylis Medical Corp *
- FDA classification date
- Nov 1, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The TorFlex Transseptal Guiding Sheath Kits are being recalled due to the possibility that the sheath could remove particulate of less than 5 mm from the dilator. The particulate could potentially migrate into a patients bloodstream.
Source facts
- Action
- Baylis Medical sent an Urgent Device Recall Notification letter dated September 23, 2013, via overnight courier to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the response form provided and fax to 905-602-5671 ATTN: Quality Department to arrange for a free of charge replacement of the affected product. The form requires consignees to return the affected product. For questions contact Quality Assurance at 800-276-4416.
- Initiated
- 2013-09-23
- Root cause
- Component design/selection
- Terminated
- 2015-03-31
- Distribution
- Worldwide Distribution - USA Nationwide in the states of : NE, NH, NC, FL, PA, VA, MI, CA, AL, and TN and the countries of China, Canada, Saudi Arabia, and Italy.
- Recall number
- Z-0085-2014
- Recall status
- Terminated
- CDRH record ID
- 122172
- FDA classified
- 2013-11-01
- Recall event number
- 66368