Skip to content

Public FDA record

Z-0085-2025: Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning t

Recall record Z-0085-2025 for LEICA BIOSYSTEMS NUSSLOCH GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0085-2025
Company in FDA record
LEICA BIOSYSTEMS NUSSLOCH GMBH
FDA classification date
Oct 15, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Source facts

Action
Leica Biosystems issued an Urgent Field Action Notice to its consignees on 09/05/2024 via mail. The notice explained the problem with the device and requested the placement of a warning sticker regarding the use of flammable freezing sprays on the device and the addition of a supplemental sheet to the current IFU. For questions, please contact: Heather Jensen, Leica Biosystems, 5205 US Highway 12, Richmond, IL 60071 USA, Tel.: +1-815-298-8371, heather.jensen@leicabiosystems.com
Initiated
2024-09-04
Root cause
Labeling design
Distribution
US Nationwide distribution.
Recall number
Z-0085-2025
Recall status
Open, Classified
CDRH record ID
209982
FDA classified
2024-10-15
Recall event number
95323

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗