Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could lead to particulate embolization, infection, thrombus or ganuloma formation from particulate matter.
Source facts
Action
On 8/18/2026, the recall notice was mailed to the customer asking them to do the following:
1) Inform individuals within your organization who need to be aware of this action.
2) Quarantine and return affected product to the firm.
4) Complete and return the Business Reply Form via email to SSSPFA@stryker.com
Questions can be directed to the firm via email, SSSPFA@stryker.com, or via phone 800-624-4422.
Initiated
2026-08-18
Root cause
Packaging process control
Distribution
US Nationwide distribution in the state of NJ.
Recall number
Z-0085-2027
Recall status
Open, Classified
CDRH record ID
222092
FDA classified
2026-10-05
Recall event number
99655
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.