The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i
Source facts
Action
A safety advisory notice, dated September 18, 2014, was sent to end users that identified the product, problem, and actions to be taken.
Initiated
2014-09-18
Root cause
Process control
Terminated
2015-02-24
Distribution
US Distribution in states of: NY, VA, and MA.
Recall number
Z-0086-2015
Recall status
Terminated
CDRH record ID
130325
FDA classified
2014-10-17
Recall event number
69371