Z-0086-2023: Veritas Advanced Infusion Packs. Used in ophthalmic procedures.

Recall record Z-0086-2023 for Johnson & Johnson Surgical Vision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0086-2023
Company in FDA record
Johnson & Johnson Surgical Vision, Inc.
FDA classification date
Oct 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

As a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.

Source facts

Action
On 09/01-02/2022, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via UPS 2-Day Mail to customers informing them that the firm is recalling specific lots of VERITAS Advanced Infusion Packs due to the potential for the irrigation luer to crack or break which could lead to reduced irrigation pressure during surgery, associated with an unstable anterior chamber. Customers are required to: 1.Identify if any of their inventory contains VERITAS Packs with a lot number listed on page 5. 2.Immediately discontinue using and remove from their inventory all affected VERITAS Packs. No other VERITAS Advanced Infusion Packs are affected by this recall. 3.Complete the attached Customer Reply Form (on page 4). We require this information for reconciliation purposes with regulatory agencies, even if you have no inventory. If customers have product to be returned: -Complete the Customer Reply Form, noting the lot numbers of the VERITAS Packs. -Contact Customer Support at 1-877-266-4543 Option 2 to obtain a RGA number and arrange the product return. -Email Customer Reply Form to RegCompliOne@its.jnj.com within 3 business days of receipt of this letter. - A credit will be issued upon receipt of the customer reply form and product. If you do not have product to be returned: Email Customer Reply Form to RegCompliOne@its.jnj.com For questions or concerns, contact 1-877-266-4543 Option 2.
Initiated
2022-09-01
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution. US states of AL, AR, AZ, CA, CO, DE, FL, GA, HI, IL IN, KY LA, MN, MO, MS, NC, NJ, NY, OH, TN, TX, UT, VA, and WA; Australia, Azerbaijan, Belgium, Brazil, Canada, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, India, Iraq, Ireland, Italy, Japan, Jordan, Kuwait, Morocco, Netherlands, New Zealand, Oman, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom
Recall number
Z-0086-2023
Recall status
Open, Classified
CDRH record ID
196039
FDA classified
2022-10-12
Recall event number
90847

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗