Open original FDA record ↗Recall record
- FDA identifier
- Z-0087-2009
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- Oct 8, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Sporadic errors in various device functions, including low readings of the single gas flows for oxygen, nitrous oxide, or air delivery, or mechanical ventilation failure. Alarms and error messages function as designed.
Source facts
- Action
- Draeger Medical sent a Medical Device Recall letter to their customers via certified mail on 8/22/08 informing them of the problem. Until the exchange of affected printed circuit boards has been implemented, the firm is recommending that users closely monitor all device displays and error messages and if applicable act in accordance with the advisory statements and remedial information provided in the Instructions for Use as well as making provisions for alternative patient care measures.
- Initiated
- 2008-08-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2009-09-01
- Distribution
- Nationwide Distribution.
- Recall number
- Z-0087-2009
- Recall status
- Terminated
- CDRH record ID
- 73333
- FDA classified
- 2008-10-08
- Recall event number
- 49442