Open original FDA record ↗Recall record
- FDA identifier
- Z-0087-2011
- Company in FDA record
- Theken Spine Llc
- FDA classification date
- Oct 19, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Manta Ray Cervical System was subject to corrective action by the firm after two (2) customer complaints were received alleging that screws were 'backing out' of the device during surgical procedures.
Source facts
- Action
- Integra sent a Technical Bulletin dated July 29, 2010 via Fedex to their customers. The bulletin identified the product, the problem, and action to be taken by the user/customers (surgeons) of the two (2) reported instances of screws backing out of the plates. The Bulletin provided instructions outlining the proper surgical techniques for implantation of screws using the Manta Ray ACP System. The bulletin also included an attached acknowledgement form which the user/customers were asked to complete and FAX to (609) 275-5363 or E-mail back to the E-mail address provided. The form verified that each customer: 1) received the technical bulletin; and 2) has read and understood the proper technique for surgical placement of the screws as outlined in the bulletin. For technical support customers were to contact the Customer Service Department at (866) 942-8698.
- Initiated
- 2010-07-29
- Root cause
- Other
- Terminated
- 2012-09-21
- Distribution
- Nationwide Distribution - USA including AL, AZ, CA, CT, FL, GA, HI, IL, ID, IN, KS, LA, MI, MD, MN, MA, MD, MS, NE, NY, NC, NJ, NV, NM, OH, OR, OK, PA, TN, TX, UT, WI, and WA.WA.
- Recall number
- Z-0087-2011
- Recall status
- Terminated
- CDRH record ID
- 93807
- FDA classified
- 2010-10-19
- Recall event number
- 56502