Z-0087-2022: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2). (Different brand names of same product.)

Recall record Z-0087-2022 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0087-2022
Company in FDA record
Inpeco S.A.
FDA record creation date
Oct 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.

Source facts

Action
On September 10, 2021, the firm distributed "Urgent IVD Medical Device Correction" letters to affected consignees. Distributors were instructed to forward the recall communication to downstream customers. Customers were informed that the potential risk associated with the issue is related to an incorrect patient result that may lead to incorrect patient treatment. Users were provided with detailed instructions on precautions to avoid the occurrence of the described issue, including a daily visual check and what to do in case of error 3160 or 0D60. Contact your local technical support provider if you need assistance implementing these actions. Your service provider will contact you to schedule the firmware upgrade of the Automation System Interface Module to ADVIA Centaur XP/XPT. Until then, please maintain awareness of this recall and follow precautions provided by the firm. If you have any questions about this recall, please contact Eva Balzarotti, Regulatory Affairs Manager at Regulatory.Affairs@inpeco.com. Phone number: (+41) 91 9118 224.
Initiated
2021-09-10
Root cause
Software design
Distribution
Domestic distribution nationwide . Foreign distribution worldwide.
Recall number
Z-0087-2022
Recall status
Open, Classified
CDRH record ID
189370
FDA record created
2021-10-12
Recall event number
88700

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗