Z-0087-2023: Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Recall record Z-0087-2023 for Shanghai United Imaging Healthcare Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0087-2023
Company in FDA record
Shanghai United Imaging Healthcare Co., Ltd.
FDA classification date
Oct 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Source facts

Action
On 5/30/22, advisory notices were emailed to customers who were asked to do the following: If the connection is not stable, perform a system reboot immediately and observe if the ECG and respiratory signals are restored. 2. If a system reboot cannot restore the wireless connection, use a third-party trigger device that supports 50ohm BNC interface and TTL output pulse with 100ms width, and connect to the system's external trigger port on the front cover. The port can be located according to Section 3.4.7 of the System Operator Manual. a) Make sure ECG and respiratory square pulse signals are displayed on the console monitor; b) Perform gated reconstruction based on the ECG and respiratory trigger pulses. 3. For further questions, please contact your recalling firm service engineer or call the Customer Contact Center's 24-hour hotline at (855) 221-1552 (toll-free) or (832) 699-2799 (direct). 4. Notify all your internal users of the PET/CT system, using this letter. 5. Complete and return the Receipt of Advisory Notification. The recalling firm will perform a design change and update the firmware of the wireless VSM module to resolve this issue. This design change is scheduled to be released for the on-site upgrade in August 2022.
Initiated
2022-05-30
Root cause
Device Design
Distribution
Distribution to US states of TX, WI, CA, and MI
Recall number
Z-0087-2023
Recall status
Open, Classified
CDRH record ID
196139
FDA classified
2022-10-13
Recall event number
90888

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗