Customer complaint during testing of module prior to production use. Software problem when the user has indicated to enter a coded comment instead of using the defined keyboard, then enters 100,000, the system willfile the number 100 instead of number 100,000.
Source facts
Action
Telephone calls and emails dated 9/29/05 to all consignees in production.
Product Safety Noticed dated 9/29/2005 was faxed to all consignees testing the product.
Initiated
2005-09-29
Root cause
Other
Terminated
2006-02-10
Distribution
Nationwide and Canada, Denmark, Emirates, Saudi Arabia, United Kingdom
Recall number
Z-0088-06
Recall status
Terminated
CDRH record ID
42358
FDA classified
2005-10-22
Recall event number
33732