Open original FDA record ↗Recall record
- FDA identifier
- Z-0088-2012
- Company in FDA record
- Go Medical Industries Pty., Ltd.
- FDA classification date
- Oct 21, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks because it was determined that the swabsticks were manufactured without having a system for microbial testing in place at the time of release.
Source facts
- Action
- The firm, Go Medical Industries Pty Ltd, sent an e-mail on September 3, 2011 and follow-up with an "URGENT: MEDICAL DEVICE RECALL" letter dated September 07, 2011 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to quarantine and not sell the product in the market place; immediately perform a sub-recall and notify their customers and complete and return the RETURN RESPONSE FORM via fax at 716-883-8099 or mail to: American Australian Medical Inc, 1083 Delaware Avenue, Buffalo, NY 14209. For further details you can contact QA Manager at Go Medical Industries at +61 89 381 3226 or pmoretta@gomedical.com.au.
- Initiated
- 2011-09-03
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-01-30
- Distribution
- Nationwide distribution.
- Recall number
- Z-0088-2012
- Recall status
- Terminated
- CDRH record ID
- 103952
- FDA classified
- 2011-10-21
- Recall event number
- 59892