- Recall classification
- Class II
Reason stated in the source
Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).
Source facts
- Action
- Distributors were sent via e-mail a SenTec 'Mandatory initiated field correction action" letter dated September 15, 2016 . The letter provided a "Description of the issue, Affected products, Identify and localize affected products, Removal from use, Replacement of affected products at end customer & distributor / sub-distributor level". It also provided "action information for sub-distributors and affiliates, and end customers". For questions contact SenTec Inc. by e-mail (customersevice.usa@sentec.com) or phone (636-343-0000).
- Initiated
- 2016-09-15
- Root cause
- Component change control
- Terminated
- 2017-06-23
- Distribution
- Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.
- Recall number
- Z-0088-2017
- Recall status
- Terminated
- CDRH record ID
- 149821
- FDA classified
- 2016-10-17
- Recall event number
- 75257