Open original FDA record ↗Recall record
- FDA identifier
- Z-0088-2027
- Company in FDA record
- Medtronic MiniMed, Inc.
- FDA classification date
- Oct 5, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
Source facts
- Action
- On 08/20/2026, Technical Support called affected customers to inform that, MiniMed identified an issue in which certain MiniMed Flex Insulin Pumps (MMT-8062) were released to customers with incomplete manufacturing testing. As a result of incomplete testing, your pump may have the potential for lack of protection from moisture exposure, slow charging, disconnection with sensor or MiniMed App, reduced ability to detect occlusion, no audio or low audio volume for alarms/alerts, or insulin under-delivery. If your pump experiences these issues, it could potentially lead to delayed therapy or risk of hypoglycemia, hyperglycemia, diabetic ketoacidosis (DKA) or death. For customers that were not able to be contacted via telephone call, an "URGENT MEDICAL DEVICE RECALL" Letter will be sent viaUSPS First-Class Mail. Customers are instructed to: 1. Stop using the Flex pump in their possession and revert to their back-up plan for insulin therapy. 2. Contact MiniMed immediately at (844) 885-0259 so arrangement can be made to ship customers a new Flex pump overnight. Upon calling, please say Flex when prompted to connect with an agent. 3. Please acknowledge that you ve read and understood this notification and have followed the actions listed by completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting [https://www.minimed.com/en-us/safety-notices/notice33-letter] to acknowledge communication. For Questions or assistance - call the MiniMed 24-Hour Technical Support line at 1-800-646-4633, option 1.
- Initiated
- 2026-08-20
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.
- Recall number
- Z-0088-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222528
- FDA classified
- 2026-10-05
- Recall event number
- 99818