Z-0088-2027: Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-807

Recall record Z-0088-2027 for Medtronic MiniMed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0088-2027
Company in FDA record
Medtronic MiniMed, Inc.
FDA classification date
Oct 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.

Source facts

Action
On 08/20/2026, Technical Support called affected customers to inform that, MiniMed identified an issue in which certain MiniMed Flex Insulin Pumps (MMT-8062) were released to customers with incomplete manufacturing testing. As a result of incomplete testing, your pump may have the potential for lack of protection from moisture exposure, slow charging, disconnection with sensor or MiniMed App, reduced ability to detect occlusion, no audio or low audio volume for alarms/alerts, or insulin under-delivery. If your pump experiences these issues, it could potentially lead to delayed therapy or risk of hypoglycemia, hyperglycemia, diabetic ketoacidosis (DKA) or death. For customers that were not able to be contacted via telephone call, an "URGENT MEDICAL DEVICE RECALL" Letter will be sent viaUSPS First-Class Mail. Customers are instructed to: 1. Stop using the Flex pump in their possession and revert to their back-up plan for insulin therapy. 2. Contact MiniMed immediately at (844) 885-0259 so arrangement can be made to ship customers a new Flex pump overnight. Upon calling, please say Flex when prompted to connect with an agent. 3. Please acknowledge that you ve read and understood this notification and have followed the actions listed by completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting [https://www.minimed.com/en-us/safety-notices/notice33-letter] to acknowledge communication. For Questions or assistance - call the MiniMed 24-Hour Technical Support line at 1-800-646-4633, option 1.
Initiated
2026-08-20
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.
Recall number
Z-0088-2027
Recall status
Open, Classified
CDRH record ID
222528
FDA classified
2026-10-05
Recall event number
99818

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗