Open original FDA record ↗Recall record
- FDA identifier
- Z-0089-2027
- Company in FDA record
- Nihon Kohden America LLC
- FDA classification date
- Oct 6, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
Source facts
- Action
- On 09/08/2026, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via USPS certified mail to customers informing that, Nihon Kohden Corporation announced the discovery of a vulnerability within a limited number of CNS-6201 models to Denial of Service (DoS) attacks. The identified vulnerability is classified as a NULL pointer dereference* (CWE-476). This type of software error may cause the system to crash when processing malformed or unexpected network input. Nihon Kohden has not received any complaints related to this vulnerability. If a CNS-6201 is affected by a DoS event, the device may stop its monitoring function due to a system crash, resulting in temporary loss of central monitoring functionality. Customers are instructed to: 1. Identify any affected CNS-6201 units in their inventory. 2. Complete the attached Medical Device Notification Return Response Form and email the completed Medical Device Notification Return Response Form back to NKA Technical Services at vitalsignsTS@nihonkohden.com. Issue Resolution a. Nihon Kohden Corporation advises that the customer upgrade to a successor product (e.g. CNS-2101). b. For customers who continue to operate CNS-6201 systems, Nihon Kohden reiterates the importance of implementing established cybersecurity controls that are part of the intended deployment architecture and long-standing cybersecurity guidance for these systems. These controls are not newly introduced mitigations but rather represent expected deployment requirements designed to support the secure operation of the system throughout its lifecycle. For questions - call Nihon Kohden America at 1-800-325-0283, Option 2 (M-F, 8am-5pm PDT) or email vitalsignsTS@nihonkohden.com.
- Initiated
- 2026-09-08
- Root cause
- Software design
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.
- Recall number
- Z-0089-2027
- Recall status
- Open, Classified
- CDRH record ID
- 221825
- FDA classified
- 2026-10-06
- Recall event number
- 99549