Z-0089-2027: Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software

Recall record Z-0089-2027 for Nihon Kohden America LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0089-2027
Company in FDA record
Nihon Kohden America LLC
FDA classification date
Oct 6, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.

Source facts

Action
On 09/08/2026, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via USPS certified mail to customers informing that, Nihon Kohden Corporation announced the discovery of a vulnerability within a limited number of CNS-6201 models to Denial of Service (DoS) attacks. The identified vulnerability is classified as a NULL pointer dereference* (CWE-476). This type of software error may cause the system to crash when processing malformed or unexpected network input. Nihon Kohden has not received any complaints related to this vulnerability. If a CNS-6201 is affected by a DoS event, the device may stop its monitoring function due to a system crash, resulting in temporary loss of central monitoring functionality. Customers are instructed to: 1. Identify any affected CNS-6201 units in their inventory. 2. Complete the attached Medical Device Notification Return Response Form and email the completed Medical Device Notification Return Response Form back to NKA Technical Services at vitalsignsTS@nihonkohden.com. Issue Resolution a. Nihon Kohden Corporation advises that the customer upgrade to a successor product (e.g. CNS-2101). b. For customers who continue to operate CNS-6201 systems, Nihon Kohden reiterates the importance of implementing established cybersecurity controls that are part of the intended deployment architecture and long-standing cybersecurity guidance for these systems. These controls are not newly introduced mitigations but rather represent expected deployment requirements designed to support the secure operation of the system throughout its lifecycle. For questions - call Nihon Kohden America at 1-800-325-0283, Option 2 (M-F, 8am-5pm PDT) or email vitalsignsTS@nihonkohden.com.
Initiated
2026-09-08
Root cause
Software design
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.
Recall number
Z-0089-2027
Recall status
Open, Classified
CDRH record ID
221825
FDA classified
2026-10-06
Recall event number
99549

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗