Open original FDA record ↗Recall record
- FDA identifier
- Z-0090-2025
- Company in FDA record
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- FDA classification date
- Oct 15, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Source facts
- Action
- Leica Biosystems issued an Urgent Field Action Notice to its consignees on 09/05/2024 via mail. The notice explained the problem with the device and requested the placement of a warning sticker regarding the use of flammable freezing sprays on the device and the addition of a supplemental sheet to the current IFU. For questions, please contact: Heather Jensen, Leica Biosystems, 5205 US Highway 12, Richmond, IL 60071 USA, Tel.: +1-815-298-8371, heather.jensen@leicabiosystems.com
- Initiated
- 2024-09-04
- Root cause
- Labeling design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0090-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210087
- FDA classified
- 2024-10-15
- Recall event number
- 95323