Audible alarm occurs before aspiration potentially resulting in operator injury or results assigned to wrong patient for STAT samples.
Source facts
Action
The firm issued a letter on 09/30/2003 notifying customers of the Device Correction and provided instructions to disable the audible alert. Abbott Diagnostics will correct this issue in the next revision of the AEROSET System Software.
Initiated
2003-09-30
Root cause
Other
Terminated
2004-04-07
Distribution
Nationwide and to the following foreign countries: Norway, United Kingdom, Hong Kong, United Arab Emirates, Canada, New Zealand, and China
Recall number
Z-0091-04
Recall status
Terminated
CDRH record ID
29968
FDA classified
2004-07-20
Recall event number
27495