Skip to content

Public FDA record

Z-0091-2021: Levo Arm Label: LEVO ARM REF 7887-050 MIZUHO OSI Refurbished Levo Arm Label: LEVO ARM REF 7887-050R MIZU

Recall record Z-0091-2021 for Mizuho OSI. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0091-2021
Company in FDA record
Mizuho OSI
FDA record creation date
Oct 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the potential for fluid ingress into the Arm which could affect the device's electrical system and lead to a failure of the locking mechanism

Source facts

Action
On 10/14/2019, the firm sent an "URGENT: Medical Device Recall" notification to customers via courier (e.g., UPS or FedEx) and an email to distributors informing them that they have become aware of the potential for fluid ingress rendering the internal electronics inoperable and leading to a failure of the Arm's locking mechanism. The Recalling Firm is instructing customers that they can continue to use the Arm until a replacement Arm is provided, but if either of the following scenarios apply to the devices, to please cease uses immediately and contact the Recalling Firm's Service Department. The two scenarios are: -The customer believes the Arm has been exposed to excessive fluids at any point while in use; -The customer notices any slipping of the Arm while properly installed for its intended use. Customers are asked to refer to the Owner's Manual for specific instructions on how to properly clean all modules of the Head Positioning System, including the Arm. Instructions are found in Section 13 of the manual. In addition, the cleaning instructions are included within the Customer Notification Letter. The Recalling Firm is proactively replacing Arms. Customer should follow the instruction regarding the Exchange Program: -Recalling Firm will provide a replacement Arm at no charge. The timing of the replacement will be based upon product availability and is expected to be completed within 16 weeks. -The Recalling Firm's Sales or Service members will assist customers with the packaging and return of the existing Arm(s) upon the receipt of the upgraded model. Customers are also asked to respond to the Customer Notification Letter by: 1. Confirm receipt and understanding of the notice by signing and returning the response portion of the Notification letter to the Recalling Firm's Regulatory Department at: FAX Number +1-510-429-9945 or email notice@mizuhosi.com 2. Any questions about the exchange plan, call the Recalling Firm's Service Department a
Initiated
2019-10-14
Root cause
Device Design
Terminated
2023-09-27
Distribution
Worldwide distribution - U.S. Nationwide distribution including in states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, SC, TN, TX, VA, WI, and WV. The countries of Australia, Belgium, Canada, Columbia, Germany, Israel, South Korea, New Zealand, Poland, United Kingdom, and South Africa.
Recall number
Z-0091-2021
Recall status
Terminated
CDRH record ID
183814
FDA record created
2020-10-09
Recall event number
86482

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗