Z-0091-2022: Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic B

Recall record Z-0091-2022 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0091-2022
Company in FDA record
Datascope Corp.
FDA classification date
Oct 20, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

Source facts

Action
On September 21, 2021, the firm distributed Urgent Medical Device Removal letters to affected customers. Customers were alerted that affected battery packs may experience unexpected short battery runtime. Actions to be Taken by Customer: - Examine your inventory to determine if you have any of the affected product - Replace any affected battery with an unaffected battery, and remove affected product from areas of use - Dispose of affected batteries properly in accordance with local statutes and the labeling on the battery pack - Customers with affected product are eligible for credit or replacement at no cost - Forward information to all current and potential Cardiosave Hybrid and Cardiosave Rescue IABP users within your hospital/facility - Distributors should forward notification to affected customers - Respond to firm's notification by completing the included response form On 11/04/2025, Datascope Getinge sent a customer update letter (i.e. Urgent Medical Device Correction) to communicate the IFU Addendum to consignees on 11/04/2025. The addendum communicates changes regarding the battery run time specification. The IFU Addendum updates the battery runtime specification to match those published in the Service Manual in June 2023. The notification also includes information regarding other issues related to the recall and being communicated to users of the Cardiosave pumps.
Initiated
2021-09-21
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium, Canada, China, Czech Republic, Egypt, France, Germany, India, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, Pakistan, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, UK, United Arab Emirates, Philippines, and Russia
Recall number
Z-0091-2022
Recall status
Open, Classified
CDRH record ID
189393
FDA classified
2021-10-20
Recall event number
88719

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗