Z-0092-2014: REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA

Recall record Z-0092-2014 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0092-2014
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA classification date
Oct 28, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma

Source facts

Action
Stryker sent an Urgent Medical Device Correction Notification letter via overnight mail and interntion consignees via E-mail on October 2, 2013. The Urgent Notice identified the recalled product, the reason for the recall, potential adverse risks to health associated with the recall. Customers were instructed to: 1. Immediately review this Recall Notification 2. Locate, and quarantine, the units listed in this notification. Do not use the recalled TPS MicroDriver until it has been repaired. A list, of units shipped to your facility, is included on the attached Business Reply Form. 3. Return the enclosed Business Reply Form to confirm receipt of this notification and identify how many affected units are currently in your inventory. 4. If you have further distributed this product, please forward this letter and the attached Business Reply Form (BRF) to all affected locations. Please indicate each location on the BRF. 5. Fax the completed Business Reply Form to Stryker Instruments Regulatory Department, 866-521-2762. 6. Upon receipt of the Business Reply Form, Stryker will send you a pre-paid shipper to send your affected device(s) back to Stryker to be repaired. Unit will be repaired and returned to your facility 7. Send back all affected devices using the pre-paid shipper provided to you by Stryker. 8. If you require a loaner, please contact the Stryker Service Department at 888-308-1983. For questions regarding this recall call 269-389-4518.
Initiated
2013-10-02
Root cause
Nonconforming Material/Component
Terminated
2014-04-15
Distribution
Worldwide distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Italy, Japan, Spain, Switzerland, and United Kingdom.
Recall number
Z-0092-2014
Recall status
Terminated
CDRH record ID
122137
FDA classified
2013-10-28
Recall event number
66352

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗