Z-0092-2021: Servo-i Ventilator, Part No. 6487800, UDI Code 07325710000823 - Product Usage: intended for general and c

Recall record Z-0092-2021 for Getinge Group Logistics America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0092-2021
Company in FDA record
Getinge Group Logistics America, LLC
FDA record creation date
Oct 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potentially shorter than specified nebulizer connector may result in difficulty in installing the Aeroneb Nebulizer module to the Servo-i device successfully.

Source facts

Action
On September __, 2020, the firm issued an Urgent Medical Device Recall letter to affected customers. Distributors were instructed to forward the information to all current and potential Servo-i Ventilator System users and biomedical staff within hospitals/facilities. Customers were asked to examine their inventory immediately to determine if they have any products with the listed serial numbers. For affected units, an alternate means of nebulization would need to be readily available and used until the unit correction can be performed by the Getinge Service Technician if nebulization via the Servo-i Aeroneb Nebulization module be required. Customers were also asked to complete and return the response form to the firm to acknowledge their receipt of the notification. If you have one or more affected systems, you will be contacted by a Getinge Service Representative to schedule on-site correction of your device as soon as an update kit is available. Affected devices will be corrected free of charge. You can also arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9438872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). You may also call this number if you have any questions about this recall.
Initiated
2020-09-17
Root cause
Device Design
Distribution
Domestic distribution to AZ CA GA IL MN MO NC NJ NY RI SC TX VA WI WV Foreign distributed to Bosnia and Herzegovina, Czech Republic, France, Germany, India, Indonesia, Iraq, Italy, Kenya, Republic of Korea, Lebanon, Libyan Arab Jamahiriya, Norway, Palestine, Peru, Qatar, Saudi Arabia,Slovakia, Sri Lanka, Sweden Thailand, Turkmenistan, Ukraine and United Arab Emirates.
Recall number
Z-0092-2021
Recall status
Open, Classified
CDRH record ID
184004
FDA record created
2020-10-09
Recall event number
86548

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗