Z-0092-2025: VasoView HemoPro Endoscopic Vessel Harvesting System, Product Codes VH-3000-W and VH-3500

Recall record Z-0092-2025 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0092-2025
Company in FDA record
Maquet Cardiovascular, LLC
FDA classification date
Oct 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There were reports of the silicone detaching from the Jaws of the Harvesting Tool during use.

Source facts

Action
An URGENT MEDICAL DEVICE- REMOVAL notification letter dated 9/20/24 was sent to customers. Actions to be taken by the customer: 1. Forward this information to all current and potential VasoView HemoPro Endoscopic Vessel Harvesting System users within your hospital/facility. 2. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. 3. Examine your inventory immediately to find and remove from use any of the VasoView HemoPro EVH Systems with the product codes and lot numbers listed in this notice. 4. Return any unused/unexpired affected product to MCV /Getinge. Contact MCV/Getinge Customer Service at 1-888-880-2874 between 6 am and 5 pm PST (Pacific Standard Time) to request a return authorization number (RMA) and shipping instructions. If you have an affected product, you are entitled to a credit. 5. Whether or not you have affected the product, please complete and sign the MEDICAL DEVICE REMOVAL - RESPONSE FORM (page 5) to acknowledge receipt of this notification. Return the completed form to MCV/Getinge by emailing a scanned copy to Hemopro-peeling-detachedsilicone2024.act@getinge.com or by faxing the form to + 1 (866) 594 8101. Actions to be taken by Getinge: MCV/Getinge is continuing to investigate this issue to determine and implement appropriate actions to prevent the recurrence of this failure. MCV /Getinge will notify customers if additional action is needed to address this issue. Getinge will facilitate the removal of affected products from your facility and provide credit for your return of these products. This removal only affects the products listed on page 1; no other products are affected by this medical device removal. If you have any questions, contact your MCV /Getinge representative or call the MCV/Getinge Customer Support at 1-888-880-2874, Monday through Friday, between 6 am and 5 pm PST.
Initiated
2024-09-20
Root cause
Device Design
Distribution
Worldwide distribution. US Nationwide. Brazil, China, and Hong Kong.
Recall number
Z-0092-2025
Recall status
Open, Classified
CDRH record ID
210264
FDA classified
2024-10-25
Recall event number
95434

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗