Open original FDA record ↗Recall record
- FDA identifier
- Z-0093-06
- Company in FDA record
- Advanced Medical Optics, Inc.
- FDA classification date
- Oct 28, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A small opening (channel or tunnel) in the seal of the outer Tyvek pouch which contains the lens case, in which the sealed intraocular lens is packaged may not be adequately sealed, therefore the potentially compromised outer Tyvek pouch may not be sterile and may potentially contaminate an otherwise sterile field during surgery.
Source facts
- Action
- A product notification letter was mailed to all US domestic customers via overnight carrier (Federal Express) beginning on October 5, 2005. A global notification letter will be translated into the local languages by the international AMO affiliates and mailed to international customers as soon as a listing of all customers is tabulated.
- Initiated
- 2005-10-05
- Root cause
- Other
- Terminated
- 2008-05-23
- Distribution
- Nationwide, Austria, Belgium, Switzerland, Germany, Denmark, France, UK, Guadelope, Hong Kong, Croatia, India, Italy, Netherlands, and Vietnam
- Recall number
- Z-0093-06
- Recall status
- Terminated
- CDRH record ID
- 42549
- FDA classified
- 2005-10-28
- Recall event number
- 33849