McKesson Cardiology Hemo front-end parameter boxes must undergo annual
maintenance testing to establish the proper functioning of the parameter boxes,
and to identify in a timely manner whether a parameter box is out of calibration.
McKesson has identified that its communication to customers regarding the
required maintenance tests was insufficient. While each unit has an attached
sticker i
Source facts
Action
Consignees were notified via Field Safety Notice on 06/25/2014.
Initiated
2014-06-25
Root cause
Labeling design
Terminated
2017-10-26
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.
Recall number
Z-0093-2015
Recall status
Terminated
CDRH record ID
130368
FDA classified
2014-10-20