Z-0093-2018: Symbia Intevo 16, SPECT/CT System, Emission Computed Tomography System To detect or image the distributio

Recall record Z-0093-2018 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0093-2018
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA record creation date
Nov 16, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Broad Quantification option of the Symbia product software version VB20A may not allow modification of two data input values. The failure occurs when the system with the Broad Quantification option is calibrated with phantoms that do not have a volume of 6500ml and/or when residual dose needs to be entered.

Source facts

Action
Siemens Healthineers sent a Customer Safety Advisory Notice Letter to each impacted consignee beginning on November 07, 2017. Depending on the customer's capabilities, Siemens provides the letters via mail, e-mail, or hand delivery by a service engineer. The receipt of the letter will be documented and additional attempts will be made until such receipt is acknowledged. Certified mail will be used to track the submission of the letter. Certified mail receipts will be obtained for each customer. Adobe e-signature will be obtained from the customers receiving the Customer Advisory Letter via email. Unsuccessful delivery will be documented and a field service engineer will be dispatched to the site to deliver the letter and applicable information. Customers were advised to please ensure that this customer advisory notice is placed in your system with the Broad Quantification option Operators Manual and disseminated to all operators of the system with the Broad Quantification option system. If this equipment is no longer in your possession, we kindly ask that you forward this letter to the new owner of the equipment, and please inform Siemens about the change in ownership. Adverse events or quality problems experienced with the use of this product should be reported to Siemens through the contact information provided below and may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you have any questions regarding this advisory notice, please contact your local Siemens representative at the contact numbers provided below. " America: 1-800-888-7436 " Europe, Middle East, and Africa: +49 9131 940 4000 " Asia and Australia: +86 (21) 3811 2121
Initiated
2017-11-07
Root cause
Software design
Terminated
2019-04-29
Distribution
Worldwide Distribution to Malaysia, France, Japan, Reunion Island
Recall number
Z-0093-2018
Recall status
Terminated
CDRH record ID
158853
FDA record created
2017-11-16
Recall event number
78184

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗