Open original FDA record ↗Recall record
- FDA identifier
- Z-0093-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- Nov 16, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Broad Quantification option of the Symbia product software version VB20A may not allow modification of two data input values. The failure occurs when the system with the Broad Quantification option is calibrated with phantoms that do not have a volume of 6500ml and/or when residual dose needs to be entered.
Source facts
- Action
- Siemens Healthineers sent a Customer Safety Advisory Notice Letter to each impacted consignee beginning on November 07, 2017. Depending on the customer's capabilities, Siemens provides the letters via mail, e-mail, or hand delivery by a service engineer. The receipt of the letter will be documented and additional attempts will be made until such receipt is acknowledged. Certified mail will be used to track the submission of the letter. Certified mail receipts will be obtained for each customer. Adobe e-signature will be obtained from the customers receiving the Customer Advisory Letter via email. Unsuccessful delivery will be documented and a field service engineer will be dispatched to the site to deliver the letter and applicable information. Customers were advised to please ensure that this customer advisory notice is placed in your system with the Broad Quantification option Operators Manual and disseminated to all operators of the system with the Broad Quantification option system. If this equipment is no longer in your possession, we kindly ask that you forward this letter to the new owner of the equipment, and please inform Siemens about the change in ownership. Adverse events or quality problems experienced with the use of this product should be reported to Siemens through the contact information provided below and may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you have any questions regarding this advisory notice, please contact your local Siemens representative at the contact numbers provided below. " America: 1-800-888-7436 " Europe, Middle East, and Africa: +49 9131 940 4000 " Asia and Australia: +86 (21) 3811 2121
- Initiated
- 2017-11-07
- Root cause
- Software design
- Terminated
- 2019-04-29
- Distribution
- Worldwide Distribution to Malaysia, France, Japan, Reunion Island
- Recall number
- Z-0093-2018
- Recall status
- Terminated
- CDRH record ID
- 158853
- FDA record created
- 2017-11-16
- Recall event number
- 78184