Z-0093-2023: N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number

Recall record Z-0093-2023 for SIEMENS HEALTHCARE DIAGNOSTICS INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0093-2023
Company in FDA record
SIEMENS HEALTHCARE DIAGNOSTICS INC
FDA classification date
Oct 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

Source facts

Action
Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction (UMDC) PP-22-003-A-C US to US customers by FedEx on 08/30/2022; and Urgent Field Safety Notice (UFSN) PP-22-003-A-C OUS -all OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 08/19/2022. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director. Discontinue use of N Antiserum to Human IgG lot 153095B for urine and CSF sample determination. However, the N Antiserum to Human IgG lot 153095B can still be used for serum sample determination. All other N Antiserum to Human IgG lots listed in Table 1 can continue to be used with the following new preliminary high dose hook thresholds for urine and CSF: " Urine: 395 mg/L " CSF: 412 mg/L Further information regarding establishment of the preliminary high dose hook threshold can be found within section Additional Information . " The specified high dose hook threshold for serum (83.3 g/L) remains unchanged for all N Antiserum to Human IgG lots. According to the IFU, in case of questionable results, the determinations should be repeated using the next higher sample dilution. For patient monitoring, consecutive immunoglobulin determinations should be performed from the same sample dilution, as far as possible. Review your inventory of the N Antiserum to Human IgG lot 153095B to determine your laboratory s replacement needs, if required, and to provide information to Siemens Healthineers for reporting to the authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Initiated
2022-08-30
Root cause
Under Investigation by firm
Distribution
Nationwide
Recall number
Z-0093-2023
Recall status
Open, Classified
CDRH record ID
196087
FDA classified
2022-10-14
Recall event number
90867

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗