Open original FDA record ↗Recall record
- FDA identifier
- Z-0094-2019
- Company in FDA record
- Widex A/S
- FDA record creation date
- Oct 10, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The reason for this field action is due to the mounting of a non-tamper-resistant (ordinary) battery drawer on some BABY440 hearing aids. The correct battery drawer is tamper-resistant and require a special tool to open it.
Source facts
- Action
- The firm, Widex, sent an "Urgent Medical Device Correction Letter" dated 8/28/18 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: a) Investigate, whether your childs BABY440 hearing aid is mounted with the correct battery drawer. Please refer to the attachment How to determine if correct battery drawer is mounted. b) If affirmative, return the product to the hearing care professional, where you bought the hearing aid, to have the tamper-resistant battery drawer mounted. If you have any questions, contact Director, Global Regulatory Affairs at 0045 44 35 57 71 or email to: krpe@widex.com.
- Initiated
- 2018-08-31
- Root cause
- Employee error
- Terminated
- 2018-12-21
- Distribution
- US Distribution to states of: GA and NY.
- Recall number
- Z-0094-2019
- Recall status
- Terminated
- CDRH record ID
- 167425
- FDA record created
- 2018-10-10
- Recall event number
- 81085