Skip to content

Public FDA record

Z-0094-2019: Widex BABY hearing aid (referred to as BABY440), powered by zinc-air batteries size 10 Intended use: BAB

Recall record Z-0094-2019 for Widex A/S. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0094-2019
Company in FDA record
Widex A/S
FDA record creation date
Oct 10, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The reason for this field action is due to the mounting of a non-tamper-resistant (ordinary) battery drawer on some BABY440 hearing aids. The correct battery drawer is tamper-resistant and require a special tool to open it.

Source facts

Action
The firm, Widex, sent an "Urgent Medical Device Correction Letter" dated 8/28/18 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: a) Investigate, whether your childs BABY440 hearing aid is mounted with the correct battery drawer. Please refer to the attachment How to determine if correct battery drawer is mounted. b) If affirmative, return the product to the hearing care professional, where you bought the hearing aid, to have the tamper-resistant battery drawer mounted. If you have any questions, contact Director, Global Regulatory Affairs at 0045 44 35 57 71 or email to: krpe@widex.com.
Initiated
2018-08-31
Root cause
Employee error
Terminated
2018-12-21
Distribution
US Distribution to states of: GA and NY.
Recall number
Z-0094-2019
Recall status
Terminated
CDRH record ID
167425
FDA record created
2018-10-10
Recall event number
81085

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗