Skip to content

Public FDA record

Z-0094-2024: Sapphire Infusion Pumps, Models: Multi-Therapy (REF 15031-000-0028), Epidural (REF 15032-000-0027) and Sa

Recall record Z-0094-2024 for EITAN MEDICAL LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0094-2024
Company in FDA record
EITAN MEDICAL LTD
FDA classification date
Oct 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Infusion Pumps with affected software revision may fail to detect air in line, which may lead to air embolism.

Source facts

Action
On 9/11/23, recall notices were emailed to customers informing them of the following: The Affected Pumps may fail to identify air in line when the following conditions are met: 1. The pump is operating on battery power (not connected to a power supply) 2. Treatments are water-like solutions (TPN and other opaque solutions are not affected) Until the Affected Pumps in your facility have been updated with new software, users are advised to only operate the effected devices as follows: i. Connect to continuous AC power during treatment (i.e., connect to a power supply) and/or; ii. Use with air eliminating filters sets: - Available Eitan Medical sets with filters: AP206-01, AP240-01, AP 214-01, AP 204-01, AP223-01, or; - Connect a 3rd party air eliminating filter set extension to the end of the Sapphire set. Alternatively, use software version - Rev15.10. Complete and return the Customer Response form via email to compliance@eitanmedical.com To report adverse events or product complaints, contact Global Complaint Management, complaints@eitanmedical.com If additional information or technical assistance, contact Customer Service and Technical Assistance, customerservice@eitanmedical.com 1-877-541-9944
Initiated
2023-09-11
Root cause
Software Design Change
Distribution
US nationwide distribution.
Recall number
Z-0094-2024
Recall status
Open, Classified
CDRH record ID
203459
FDA classified
2023-10-20
Recall event number
93105

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗