- FDA identifier
- Z-0094-2024
- Company in FDA record
- EITAN MEDICAL LTD
- FDA classification date
- Oct 20, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Infusion Pumps with affected software revision may fail to detect air in line, which may lead to air embolism.
Source facts
- Action
- On 9/11/23, recall notices were emailed to customers informing them of the following: The Affected Pumps may fail to identify air in line when the following conditions are met: 1. The pump is operating on battery power (not connected to a power supply) 2. Treatments are water-like solutions (TPN and other opaque solutions are not affected) Until the Affected Pumps in your facility have been updated with new software, users are advised to only operate the effected devices as follows: i. Connect to continuous AC power during treatment (i.e., connect to a power supply) and/or; ii. Use with air eliminating filters sets: - Available Eitan Medical sets with filters: AP206-01, AP240-01, AP 214-01, AP 204-01, AP223-01, or; - Connect a 3rd party air eliminating filter set extension to the end of the Sapphire set. Alternatively, use software version - Rev15.10. Complete and return the Customer Response form via email to compliance@eitanmedical.com To report adverse events or product complaints, contact Global Complaint Management, complaints@eitanmedical.com If additional information or technical assistance, contact Customer Service and Technical Assistance, customerservice@eitanmedical.com 1-877-541-9944
- Initiated
- 2023-09-11
- Root cause
- Software Design Change
- Distribution
- US nationwide distribution.
- Recall number
- Z-0094-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203459
- FDA classified
- 2023-10-20
- Recall event number
- 93105