Skip to content

Public FDA record

Z-0095-2018: iBalance(R) UKA, Femoral Cemented, Size 3, Left Medial/Right LATERAL, REF AR 501-UFLC, QTY 1, Rx ONLY, ST

Recall record Z-0095-2018 for Arthrex, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0095-2018
Company in FDA record
Arthrex, Inc.
FDA record creation date
Nov 16, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Arthrex part number AR-501-UFLC, size 3, batch 10123297, has the incorrect part number AR-501 UFRA, size 1 printed on the patient label inside the box.

Source facts

Action
Arthrex sent an Urgent Medical Device Voluntary Recall letter on October 3, 2017. The letter requested the following: "Immediately discontinue use of these devices. In order to receive credit for the returned devices, please follow the steps below: 1) Review this notification and ensure affected personnel are aware of the contents. 2) Immediately identify and return all the indicated product/batch numbers you have in inventory. Call Arthrex Field Logistics Service at 1-866612-7899 for a returned goods authorization (RGA) and product return instructions. Email: FLS@Arthrex.com." Customers with questions were instructed to call Product Surveillance at 866-267-9138 or by email to commplaints@arthrex.com.
Initiated
2017-10-03
Root cause
Process control
Terminated
2018-12-04
Distribution
Worldwide Distribution - US including SD, NV, NC, CA, MD, AZ, CA, LA, FL, PA, TX, NY, GA, and Internationally to GERMANY, POLAND
Recall number
Z-0095-2018
Recall status
Terminated
CDRH record ID
159082
FDA record created
2017-11-16
Recall event number
78297

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗