Skip to content

Public FDA record

Z-0095-2024: MED-810A Zemits NDPrime Laser products

Recall record Z-0095-2024 for Advance-Esthetic LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0095-2024
Company in FDA record
Advance-Esthetic LLC
FDA classification date
Oct 13, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.

Source facts

Action
Notification letters will be sent to affected customers identifying the reason for notification and action to be taken. The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For any queries or require further assistance, our customer support team is available at 888-999-3996 or customer-support@advance-esthetic.us.
Initiated
2023-10-09
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-0095-2024
Recall status
Open, Classified
CDRH record ID
203630
FDA classified
2023-10-13
Recall event number
93194

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗