Z-0096-2021: SwiftHook sold as the following: a. Individually as SwiftHook b. Included with Vega505EE mobile lifts Pro

Recall record Z-0096-2021 for Handicare AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0096-2021
Company in FDA record
Handicare AB
FDA record creation date
Oct 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Carry bar may detach during use of mobile lifts and ceiling lifts equipped with the Swifthook component.

Source facts

Action
Handicare USA issued an field safety notification on August 21, 2020. It states the following. Users of affected units are urgently requested to perform the following actions for devices in their possession without delay upon receiving the field notice. It is advised to cease using the Swifthook. Instead, the carry bar can be mounted directly onto the mobile or ceiling lift. This action will eliminate the possibility of incorrectly mounting the Swifthook while gaining the advantage of increased lift height capacity. The Swifthook should be destroyed. 1. Remove the Swifthook from the carry bar using the appropriate Allen key (5 mm) and wrench (13 mm); 2. Remove the fasteners from the mobile lift's boom using two 13 mm wrenches. In the case of a ceiling lift, remove the fasteners from the ceiling lift's H-adapter using two 4 mm Allen Keys; 3. Attach the carry bar directly to the mobile lift's boom. Check that the lock nuts are undamaged and still offers resistance to tightening by hand, and if not, replace with new nuts (DIN985 M8fzb lock nut). In the case of a ceiling lift, use two 4 mm Allen Keys and tighten well by hand. Each customer is required to communicate the content of the field safety notice to all end users and complete the response form, which shall be returned by email to Handicare USA using provided contact details no later than September 28, 2020. For any questions regarding this FSN, please contact FieldAction.NorthAmerica@handicare.com or Laren Mcintosh, Director of Quality, North America.
Initiated
2020-08-21
Root cause
Use error
Terminated
2023-01-30
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, IL, MD, MI, MN, NH, NJ, NY, ON, PA, TX, VA, WA, WI and the countries of Australia, Belgium, Canada, Chile, Denmark, Finland, Germany, Iceland, Ireland, Israel, Japan, Lithuania/ Lathvia, Malaysia, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Spain, Sweden, Taiwan the Netherlands, UAE, UK.
Recall number
Z-0096-2021
Recall status
Terminated
CDRH record ID
183334
FDA record created
2020-10-09
Recall event number
86318

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗