Skip to content

Public FDA record

Z-0097-2019: Codan Filter (Codan catalog number BC 693) used with Medrad Patient Administration Sets (PAS) and Intego

Recall record Z-0097-2019 for Bayer Medical Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0097-2019
Company in FDA record
Bayer Medical Care, Inc.
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bayer has received reports of loose particulate in the filter when the dust cap is removed. Bayer conducted further testing confirming the dust cap may generate particulate which may potentially enter the fluid path.

Source facts

Action
Urgent Medical Device Recall notification letters dated 9/7/18 were distributed to customers. The letters instruct customers to do the following: 1. Immediately discontinue use of your Intego system and quarantine the unused PAS product until further notice. 2. Contact Bayer Customer Care at 1-800-633-7231 opt. 2 or ricustomercare@bayer.com to receive a Return Goods Authorization (RGA) number and return the Codan Filters to Bayer accordingly. 3. Complete the response form indicating if your institution does or does not have affected product. Return the form via fax (1-412-406-0942) or email (randicommercialquality@bayer.com).
Initiated
2018-09-07
Root cause
Nonconforming Material/Component
Terminated
2019-07-11
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, and WI. Foreign countries: UAE, Austria, Australia, Bosnia and Herzegovina, Belgium, Bahrain, Canada, China, Chile, Costa Rica, Czechia, Denmark, German, Spain, Estonia, Egypt, Finland, France, Great Britain, Croatia, Hungary, Israel, Italy, Jordan, Kuwait, Kazakhstan, Lebanon, Lithuania, Morocco, Malta, Mexico, Netherlands, Norway, Oman, Panama, Poland, Portugal, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, and Turkey
Recall number
Z-0097-2019
Recall status
Terminated
CDRH record ID
167072
FDA record created
2018-10-11
Recall event number
80945

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗