Open original FDA record ↗Recall record
- FDA identifier
- Z-0098-2021
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA record creation date
- Oct 13, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.
Source facts
- Action
- Fresenius Medical Care issued letter dated 9/1/20 stating reason for recall, health risk and action to take: A Fresenius Medical Care Renal Therapies Group Water Systems Service Specialist will be contacting you to schedule an appointment to check the connection and replace if necessary. In the meantime, we ask that you do not perform any repairs or maintenance on the machine and wait for your scheduled service date. For questions or concerns or if you need service prior to your scheduled service date, please contact Technical Support Services at 888-553-1109.
- Initiated
- 2020-09-01
- Root cause
- Process control
- Terminated
- 2021-01-20
- Distribution
- US distribution to DE, NC, OK, PA, and TX.
- Recall number
- Z-0098-2021
- Recall status
- Terminated
- CDRH record ID
- 183655
- FDA record created
- 2020-10-13
- Recall event number
- 86410