Skip to content

Public FDA record

Z-0098-2024: REVERSE PROSTHESIS SHOULDER SYSTEM, SCREW, LOCKING BONE, RSP, 5mmX30mm long, REF: 506-03-130

Recall record Z-0098-2024 for Encore Medical, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0098-2024
Company in FDA record
Encore Medical, LP
FDA classification date
Oct 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reverse shoulder prosthesis package that should contain 5mmx30 screws could contain incorrect 3.5mmx30 screws. Use of incorrect screws could result in surgical delay and inadequate fixation.

Source facts

Action
On 8/22/23, recall notices were emailed to agents and registration holders who were asked to do the following: Contact Customer Service at 1-800-456-8696 to assist with return and replacement. This notice needs to be passed on to all those within your organization or to any organization where the potentially affected product has been transferred. Complete and return the acknowledgement and response form. If you have questions, please email productsafety@enovis.com
Initiated
2023-08-22
Root cause
Mixed-up of materials/components
Distribution
US: FL, IN, WI, IA, MN, KY, NY, AZ, RI, WA, OH, SC, VA, TX. OUS: MX
Recall number
Z-0098-2024
Recall status
Open, Classified
CDRH record ID
203401
FDA classified
2023-10-12
Recall event number
93080

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗