Philips Medical Systems Sales & Service Region No. America
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
The cable carrier and cables may fall after repeat collisions with OR lamps, radiation shields or other supported accessories.
Source facts
Action
The firm sent a letter dated, October 10, 2003 indicating the firm''s service representatives will upgrade the cable carrier by mounting a safety catch inside the cable carrier.
Initiated
2003-10-10
Root cause
Other
Terminated
2005-01-13
Distribution
The firm distributes to hospitals and medical centers nationwide.
Recall number
Z-0099-04
Recall status
Terminated
CDRH record ID
29939
FDA classified
2004-07-20
Recall event number
27485
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.