Open original FDA record ↗Recall record
- FDA identifier
- Z-0099-2021
- Company in FDA record
- Gentian AS
- FDA record creation date
- Oct 13, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has identified lots of calibrators that have shown to measure incorrect high values over time compared to the assigned values when released for distribution. This could result in higher Cystatin C values measured in patients.
Source facts
- Action
- Gentian AS issued a "Field Safety Notice" letter on September 2, 2020 concerning the removal of the Gentian Cystatin C Calibrator and Calibrator Kit. In addition, to informing consignees about the recall the firm asked consignees to take the following actions: 1. Discontinue use of the affected lots listed in this Field Safety Notice 2. Please inform Gentian of remaining stock of the affected lots numbers given in this Field Safety Notice. 3. Share this notification with other relevant personnel in your organization and affected customers 4. Please sign and return this document by e-mail to confirm that this notification of withdrawal has been received and understood.
- Initiated
- 2020-09-02
- Root cause
- Under Investigation by firm
- Terminated
- 2022-09-28
- Distribution
- US: MN, VA, KS,
- Recall number
- Z-0099-2021
- Recall status
- Terminated
- CDRH record ID
- 183752
- FDA record created
- 2020-10-13
- Recall event number
- 86465