Open original FDA record ↗Recall record
- FDA identifier
- Z-0099-2024
- Company in FDA record
- Alcon Research, LLC
- FDA classification date
- Oct 11, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.
Source facts
- Action
- Starting on 8/18/23, Field Service Engineers told customers their devices needed pneumatics module replacements, and this replacement activity was subsequently completed. On 9/29/23, correction notices were mailed to customers informing them vision systems were corrected and asking they complete and return the response form to Market.Actions@Alcon.com Should you have any questions or concerns call Customer Service at 1-800-862-5266.
- Initiated
- 2023-08-18
- Root cause
- Component change control
- Distribution
- US: GA, IL, PR
- Recall number
- Z-0099-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203275
- FDA classified
- 2023-10-11
- Recall event number
- 93010