Z-0100-2024: Access Thyroglobulin Calibrators

Recall record Z-0100-2024 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0100-2024
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Oct 11, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.

Source facts

Action
Beckman Coulter issued an Urgent Medical Device Recall letter dated 06/28/2023 to its consignees affected worldwide on 07/06/2023 via email or letter and followed with an updated letter (dated 07/07/2023) issued on 07/14/2023 which contained FDA suggested edits. The notice explained the problem, the risk to the patient and requested the following actions be taken: If you have downloaded the alternative calibrator card issued in April 2023, please discard that calibrator card, and do not use it moving forward. . Only use Access Thyroglobulin calibrator lots 338013 or higher to calibrate Access Thyg reagent packs (C71762), when this becomes available in your laboratory. . Continue to use any lot of Thyroglobulin Calibrator (PN 33865) with Access Thyroglobulin Reagent packs (33860). The firm issued an Important Product Notification letter was issued on 2023 August 10 notifying customers in the US, Canada, and China that they were NOT affect by this issue and to disregard the prior Urgent Medical Device Recall letters.
Initiated
2023-07-06
Root cause
Other
Distribution
worldwide except US, Canada, and China
Recall number
Z-0100-2024
Recall status
Open, Classified
CDRH record ID
202089
FDA classified
2023-10-11
Recall event number
92670

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗