Reason stated in the source
The LCD of the OneTouch SureStep Meter may display incorrectly, resulting in missing segments on the display. This display failure may result in patients misreading the blood glucose result, or cause a delay in test interpretation, and could cause delays or mistreatment.
Source facts
- Action
- On 9/9/05, ***and 4/14/06 (expanded)***the firm notified its consumers, Health Care Professionals (HCPs), pharmacies, and direct accounts (including wholesalers and distributors) by letter and phone calls.
- Initiated
- 2005-09-09
- Root cause
- Other
- Terminated
- 2006-12-18
- Distribution
- Total amount of product distributed in the U.S. is 19,177 units, to distribution centers to be distributed to retailers, Health Care Professionals, wholesalers, and mail order companies. The product was distributed throughout the United States; 29,377 were distributed to Asia, Europe and Latin America.