Open original FDA record ↗Recall record
- FDA identifier
- Z-0101-2008
- Company in FDA record
- Steris Corporation Hopkins Facility
- FDA classification date
- Oct 24, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Loose or missing adaptor: The center insert of the adaptor #6 restrictor could become loose and/or come out of the adaptor's housing. The absence of adaptor #6's center insert may affect the sterilization process of the attached endoscope.
Source facts
- Action
- STERIS has contacted affected direct accounts by phone to notify direct accounts of the recall. On 9/28/2007, STERIS will mail a recall notification to all affected direct accounts via certified mail. The notification informs direct accounts that STERIS representatives will be visiting direct accounts to retrieve affected products and provide replacements. Where direct accounts further distribute, STERIS representatives will inform the direct accounts to contact their sub-accounts and facilitate exchange of affected product for replacement product. Replacement product will be available beginning November 1, 2007.
- Initiated
- 2007-09-12
- Root cause
- Component design/selection
- Terminated
- 2010-09-28
- Distribution
- Worldwide: United States, Australia, Bahrain, Canada, Germany, Great Britain, Hungary, Italy, Qatar, and Spain
- Recall number
- Z-0101-2008
- Recall status
- Terminated
- CDRH record ID
- 64407
- FDA classified
- 2007-10-24
- Recall event number
- 44810