Open original FDA record ↗Recall record
- FDA identifier
- Z-0101-2015
- Company in FDA record
- Estill Medical Technologies, Inc
- FDA classification date
- Oct 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t
Source facts
- Action
- Thermal Angel sent a Notice of Corrections letter dated July 25, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to communicate the contents of the notice to all employees, consignees, consultants, or other parties affected by this correction. The firm also updated the labeling and instructions for use. For additional information cusotmers were instructed to contact Medical Technologies at (877)354-0286.The recalling firm issued a Notice of Corrections to its distributors and customers. For questions regarding this recall call 214-561-6001.
- Initiated
- 2014-07-25
- Root cause
- Device Design
- Terminated
- 2015-08-21
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Serbia.
- Recall number
- Z-0101-2015
- Recall status
- Terminated
- CDRH record ID
- 130149
- FDA classified
- 2014-10-21
- Recall event number
- 69242