Open original FDA record ↗Recall record
- FDA identifier
- Z-0101-2018
- Company in FDA record
- ICU Medical Inc
- FDA record creation date
- Nov 16, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
(1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the battery is discharged or is disconnected; and (2) when the user accesses the Preventive Maintenance Screen in Service/Biomed Mode with a Total Delivery Time >1,500 hours, the user will not be able to interact with the device and the device must be power cycled.
Source facts
- Action
- The recalling firm issued letters dated 10/30/2017 via UPS on 10/30/2017 notifying their customers of the two issues.
- Initiated
- 2017-10-30
- Root cause
- Under Investigation by firm
- Terminated
- 2019-08-28
- Distribution
- Distribution was nationwide to medical facilities. Foreign distribution was made to Canada and Australia. Government and military distribution was also made.
- Recall number
- Z-0101-2018
- Recall status
- Terminated
- CDRH record ID
- 159159
- FDA record created
- 2017-11-16
- Recall event number
- 78312