Z-0101-2019: Enterprise Imaging Desktops Enterprise Imaging Desktops (Radiology Suite/Cardiology Suite) when deployed

Recall record Z-0101-2019 for Agfa Healthcare NV. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0101-2019
Company in FDA record
Agfa Healthcare NV
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Adding an additional comparison study to the Clinical Side Bar in the Enterprise Imaging Diagnostics Desktop, can result in a change of the sort order of the studies in the Clinical Side Bar. The thumbnail images of the studies are no longer chronologically sorted with the most recent study at the top.

Source facts

Action
The firm, Agfa, sent an "URGENT FIELD SAFETY NOTICE Medical Device Recall" Notice dated 8/3/18 to its customers. The notification described the product, problem and actions to be taken. The notification letters instructed customers to do the following: 1. Please instruct your users to pay extra attention when selecting a study in the Clinical Side Bar to be sure they are reporting on the active study and not on one of the comparison studies. The active and comparison study are clearly differentiated from each other via blue (active) and grey (comparison) icons in both the Text and Image Area. The images show the date the study was performed. 2. We recommend that an upgrade be applied to correct this issue. Refer to the table for affected versions and associated correction. An Agfa HealthCare service representative will contact you to arrange for a date to deploy the appropriate upgrade at your facility. Please distribute this information within your facility to all those who need to be aware of it. If you have any questions about this matter, please contact your local Agfa HealthCare organization 1-866-599-3868 and please reference Problem Record PRB0056200 and Vigilance Record VR0000568.
Initiated
2018-08-03
Root cause
Software design
Terminated
2020-04-03
Distribution
US distribution to states of: CA, FL, IN, LA, MN, PA, SC, TX, VA, and Puerto Rico and internationally to: Canada.
Recall number
Z-0101-2019
Recall status
Terminated
CDRH record ID
167192
FDA record created
2018-10-11
Recall event number
80998

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗