Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis.
Source facts
Action
Fresenius Medical Care issued an Urgent FMCNA LIBERTY Cycler Set Recall letter on September 3, 2010 identifying the specific affected part numbers and product lots and requesting that customers identify and discontinue the use of the affected lots immediately. Customers were requested to return a receipt of notification. Customers may arrange return unused product to FMCNA by contacting customer service at 1-800-323-5188.
Initiated
2010-09-03
Root cause
Nonconforming Material/Component
Terminated
2023-02-16
Distribution
Nationwide Distribution: Throughout USA.
Recall number
Z-0103-2011
Recall status
Terminated
CDRH record ID
94220
FDA classified
2010-10-20
Recall event number
56675