Z-0103-2019: Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and P

Recall record Z-0103-2019 for Thoratec Switzerland GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0103-2019
Company in FDA record
Thoratec Switzerland GMBH
FDA classification date
Oct 12, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.

Source facts

Action
On 09/05/2018, the firm mailed or hand delivered Urgent Medical Device Correction notices to customers informing them of the following: 1) A full backup system must be in the immediate vicinity of every patient supported by the circulatory support system, 2) If the circulatory support system motor stops, exchange the pump to the backup system as described in the circulatory support system Operating Manual. Improperly exchanging only the motor to a backup console could cause damage or harm, and 3) Inspection of the circulatory support system motor, including the cable, for damage should be conducted prior to use. Do not use if damaged. If the cable is damaged and not visible as part of the inspection, the console will alarm during the procedure. Firm provided customers with updated labeling and instructions for use. Firm informed customers that changes are being qualified to make the motor cable more resistant to damage and the they would inform customers when these designs become available. In addition, customers were asked to complete and return the acknowledgement form to the firm. Customers with questions are encouraged to call 1-800-456-1477; backup systems can also be rented by calling this number.
Initiated
2018-09-05
Root cause
Device Design
Terminated
2021-04-28
Distribution
U.S.: OR, AZ, CA, ID, WA, WI, NY, MA, CT, NJ, ME, RI, IL, MI, TN, OH, IN, KY, IA, AL, FL, NC, GA, SC, VA, MD, DC, PA, WV, DE, MN, MT, NE, CO, NM, UT, AR, MO, MC, TX, OK, LA, MS, ND
Recall number
Z-0103-2019
Recall status
Terminated
CDRH record ID
166940
FDA classified
2018-10-12
Recall event number
80917

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗