Skip to content

Public FDA record

Z-0103-2023: Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular proc

Recall record Z-0103-2023 for BALT USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0103-2023
Company in FDA record
BALT USA, LLC
FDA classification date
Oct 20, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product pouch label does not match up with carton label

Source facts

Action
On 09/07/2022, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via priority mail to customer informing them that Balt USA was informed that one device had information on the pouch label that differed from the information on the carton label. This hazard presents a high risk to patient safety with the possibility that the user could inadvertently select the wrong size coil for a procedure. Distributor/Subsidiaries are instructed to: -Inform customers about this notice. -Return to Balt USA products from Lot #F2206001068 and Lot #F220601089 Direct Customers are instructed to: -Inform within their hospital Safety Office, Pharmacists, Head of Neuroradiology and staff of Recall. -Return to Balt USA products from Lot #F2206001068 and Lot #F220601089 For questions or assistance - contact Balt USA Quality Department at email - QA@baltgroup.com telephone number - 949-788-1443 or Fax 949-788-1444
Initiated
2022-09-07
Root cause
Error in labeling
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of FL, GA, NC, and PA. The countries of Dominican Republic, India and Japan.
Recall number
Z-0103-2023
Recall status
Open, Classified
CDRH record ID
196100
FDA classified
2022-10-20
Recall event number
90870

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗