Z-0104-2009: Zyoptix XP Microkeratome tray, Ref. #ZXP1000, which contains a number of components, one of which is the

Recall record Z-0104-2009 for Bausch & Lomb Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0104-2009
Company in FDA record
Bausch & Lomb Inc
FDA classification date
Oct 22, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The plastic tip on the drive shaft may come off the shaft and cause the blade to stop oscillating.

Source facts

Action
Phone calls were initiated on 4/28/08 and an Urgent Recall letter dated 4/23/08 was issued 4/28/08 via overnight mail to U.S. customers explaining the reason for recall. The letter also requested that customers stop using the motors identified and return of the motors for evaluation. A return form was enclosed.
Initiated
2008-04-23
Root cause
Device Design
Terminated
2012-04-02
Distribution
Worldwide Distribution --- including USA and countries of Brazil, The Netherlands, Canada, India, Malaysia, Mexico, and Singapore.
Recall number
Z-0104-2009
Recall status
Terminated
CDRH record ID
72813
FDA classified
2008-10-22
Recall event number
47839

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗