Z-0104-2018: HARVEST(R) TERUMOBCT, BMAC2-120-01Bone Marrow Procedure Pack, REF 51423, Rx Only, STERILE EO

Recall record Z-0104-2018 for Terumo BCT, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0104-2018
Company in FDA record
Terumo BCT, Inc.
FDA record creation date
Nov 16, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The patient labels included in the BMAC system procedure packs, catalog numbers 51416 and 51423, are printed with an incorrect lot number, catalog number, and expiry date; the lot number on the procedure pack box label is correct. The expiration noted was not beyond initiation date of the recall. Since the product is noted as sterile, the clearance for K103340, K052925 and K991430 raises question if product was cleared with a sterility claim. Further post-market follow-up will be required to assess the safety risk of the product.

Source facts

Action
The firm initiated their recall by letter on 10/20/2017. The letter stated the following: "ACTIONS REQUIRED FOR HEALTHCARE PROVIDERS AND DISTRIBUTORS 1. For product shelved in the original procedure pack (white box), examine the label for catalog/lot numbers as shown in Figure C above. 2. If affected product is in your inventory, contact the Terumo BCT Customer Support Center at +1.877.3.FYI.BCT (U.S. Toll-Free +1.877.339.4228) or +1.303.231.HELP (+1.303.231.4357), Terumo BCT Europe N.V. at +32.2.715.0590, or your local Terumo BCT representative, and Terumo BCT will issue a Return Goods Authorization for product return. 3. Continue to use your unaffected product in accordance with the instructions for use. 4. Distribute this notification to all Harvest Bone Marrow Aspirate Concentrate (BMAC) System users within your organization. 5. IMPORTANT: Complete the attached acknowledgement and return it by fax or email to Terumo BCT by 27 October 2017. Your return of the acknowledgement is critical so we can confirm that you have received the recall notice.
Initiated
2017-10-20
Root cause
Process control
Terminated
2020-08-21
Distribution
CA, Co, MO, NY, PA, TX, VA, WI, Turkey, Austria, Spain
Recall number
Z-0104-2018
Recall status
Terminated
CDRH record ID
159144
FDA record created
2017-11-16
Recall event number
78307

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗