Open original FDA record ↗Recall record
- FDA identifier
- Z-0105-06
- Company in FDA record
- General Electric Med Systems LLC
- FDA classification date
- Oct 28, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Cracked welds on the main rotation shaft could potentially lead to the collapse of the gantry yokes that support the detectors if the system is continued to be used after the weld has completely fractured.
Source facts
- Action
- GE Healthcare issued an Urgent Safety Notice, dated 07/22/05, to all affected sites. This letter describes the potential hazard, its waring signs and the additional corrective actions that will be taken by GE Healthcare.
- Initiated
- 2005-09-27
- Root cause
- Other
- Terminated
- 2006-12-26
- Distribution
- Within the US to include: AL, AZ, CA, DE, FL, GA, IA, IL, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TX, VA, WI and WV. OUS to include: Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Greece, Italy, Korea, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sinapore, Spain, Sweden and United Kingdom
- Recall number
- Z-0105-06
- Recall status
- Terminated
- CDRH record ID
- 42516
- FDA classified
- 2005-10-28
- Recall event number
- 33832