The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in an inability to remove secretions and cause partial or complete obstruction of the airway.
Source facts
Action
The firm's sales representative was notified of the recall on September 22, 2009 via email and the sole customer was notified by letter dated September 23, 2009. All product was removed from the customer account on September 28, 2009 and returned to the recalling firm.
Initiated
2009-09-22
Root cause
Process control
Terminated
2010-08-03
Distribution
Illinois.
Recall number
Z-0105-2010
Recall status
Terminated
CDRH record ID
85826
FDA classified
2009-11-12
Recall event number
53550