Z-0106-2011: Varian brand TrueBeam and TrueBeam STx, Model Number:TMX- H19, Distributed by and Manufactured by: Varia

Recall record Z-0106-2011 for Varian Medical Systems Oncology Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0106-2011
Company in FDA record
Varian Medical Systems Oncology Systems
FDA classification date
Oct 20, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Due to a software anomaly, guidance-based couch shift values may not be applied as expected when the operator selects "Apply Shift" and presses Motion Enable buttons. This can result in the actual patient position differing from that position indicated by the imaging application affecting poistion accuracy and precision.

Source facts

Action
Varian Medical Systems issued an Urgent Medical Device Correction letter date dated May 24, 2010 to users identifying the affected device, the issue prompting the correction, and the actions to be taken by the user. Customers were asked to set couch tolerance values to small numbers, and for radiosurgical treatment tolerance values to be set to zero. A Varian representative will contact customers to schedule installation of a software version correcting the issue. US Customers can contact Varian at 1-888-827-4265,. European customers can contact Varian at +41 41 749 8844.
Initiated
2010-05-24
Root cause
Software design
Terminated
2011-07-20
Distribution
Worldwide Distribution: Two consignees, one in the US and one in Switzerland.
Recall number
Z-0106-2011
Recall status
Terminated
CDRH record ID
93278
FDA classified
2010-10-20
Recall event number
56346

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗